1. First, complete an ANSI [Linear] or [AGC] test.
2. Click the right mouse button to raise the home menu, then select [Telecoil].
3. Set the aid to "T" (telecoil input).
4. Click [Start test loop test]. Setup instructions will be displayed.
5. Position the hearing instrument on the cross on the test box floor vertically oriented as worn by a seated
individual as shown.
6. Without moving the hearing instrument, press to run the SPL in a vertical magnetic field (SPLIV)
curve or press to abort the test.
If an RTG has been previously obtained, the RTLS will be shown.